Answer: How to Operate Drive Compressor Nebulizer?Before each use inspect the Compressor: https://www.healthproductsforyou.com/c-compressors.html and nebulizer cup assembly for damage or wear, replace if required.
Place the Compressor on a table or other flat stable surface. Be sure you can easily reach the controls when seated. Do not use this device on the floor.
With the power switch in the O (OFF) position, plug the power cord into an appropriate electrical wall outlet.
Connect one end of the tubing to the compressor air-outlet connector.
Assemble the nebulizer cup and add indicated medication to the nebulizer’s cup before use.
Attach the other end of the tubing into the air-inlet connector found at the bottom of the nebulizer cup.
Turn on the Compressor by pressing the power switch to the I (ON) position and begin treatment.
If treatment needs to be interrupted, simply press power switch to O (OFF) position
When the treatment is complete, turn off the compressor by pressing the power switch to O (OFF) position and unplug the unit from the electrical outlet.
Always unplug this device immediately after use.
DO NOT use this product in or near explosive environments, or where aerosol (spray) products are being used.
DO NOT pump combustible liquids or vapors with the product or use in or near an area where flammable or explosive liquids or vapors may exist.
DO NOT operate this product in an oxygen enriched environment, i.e. oxygen tent or oxygen hood.
Use only in well ventilated areas.
CAUTION:The Compressor is designed for intermittent use only. Do not operate it continuously for more than 20 minutes for a single use without turning it off and following a cooling period for least 40 minutes.
Medical Supplies Club Customer Care
Store answer