Answer: Cleaning Instructions: Scrub the probe tip and head with a soft bristle brush and detergent.
Rinse the probe head thoroughly with clean, warm water and dry with a clean towel (air dry is permissible).
Wipe the cable with either chlorine bleach (sodium hypochlorite, 2%), isopropanol (70%), hydrogen peroxide (3%) or activated glutaraldehyde (Cidex™ Plus).
Disinfection Instructions:
After cleaning the probe tip, head and cable, soak the probe head in:
Sodium hypochlorite (2%) for fifteen minutes or,
Hydrogen peroxide (3%) for fifteen minutes or,
Activated glutaraldehyde (Cidex™ Plus) for 20 minutes or,
Isopropanol (70%) for 30 minutes.
Medical Supplies Club Customer Care
Store answer
Helpful?